next up previous
Next: About this document

Name:
April 1, 1997
Instructor: C. Davila

  1. Determine tex2html_wrap_inline31 and tex2html_wrap_inline33 for each each of the two defibrillation pulses shown below assuming each pulse is to deliver 300J to a tex2html_wrap_inline35 load. Which of the two pulses is more desirable in terms of minimizing tissue damage due to heating? Explain your reasoning.

    figure10

  2. The hypothetical ECG pulse shown was measured from a patient having an implanted DDD pacemaker. One of the atrial beats and one of the ventricular beats resulted from stimulation by the pacemaker, the other beats occurred naturally. Label the ECG components associated with the pacemaker generated atrial and ventricular beats, then draw a diagram of the DDD pacemaker and explain your reasoning.

    figure14

  3. Assume that you have access to the signals: tex2html_wrap_inline37 and tex2html_wrap_inline39 . Assume that the highest frequency components of tex2html_wrap_inline41 and tex2html_wrap_inline43 are well below tex2html_wrap_inline45 . Design a system for recovering tex2html_wrap_inline43 using the fewest number of parts (multipliers, adders, filters, etc.). Justify that your design is correct. Briefly describe a biomedical application of your design.
  4. A Fleisch pneumotachometer's calibration curve was measured in the lab and is plotted below:

    figure17

    Develop a mathematical model,

    displaymath49

    for the pneumotachometer, where tex2html_wrap_inline51 is the pressure drop across the capillary tubes and Q is flow. What is the useful operating range in Hz for the pneumotachometer? Explain your answer.





Carlos Davila
Tue Apr 1 11:53:30 CST 1997